Fuel and Lube Oil Test Cabinets

The Fuel and Lube Oil Test Cabinet is a comprehensive, portable laboratory designed for on-site monitoring of fuel, lubricating, and hydraulic oil condition. It allows engineers to perform key analyses quickly and accurately, helping to detect contamination, wear, and degradation before damage occurs. Immediate results enable informed maintenance decisions, improving reliability and efficiency across all vessel systems.

For more detailed information on this product including a comparison between the versions, please view our data sheet.

Features

  • Comprehensive Testing: Includes viscosity, water, base number, and contamination checks.
  • Portable Design: Compact and mountable for use onboard or in workshops.
  • Custom Configurations: Available in basic or advanced versions with additional test options.
  • Efficient Operation: Built-in lighting and workspace for convenient testing.

Benefits

  • Accurate Diagnostics: Detects early signs of oil degradation and contamination.
  • Reduces Downtime: Enables proactive maintenance and quicker response.
  • Flexible Options: Customisable for specific vessel or oil type requirements.
  • Improves Reliability: Extends equipment life through regular monitoring.

Products in this range

Show

entries

Previous Next
Test Measuring Range Measuring Time Accuracy / Notes

Test

Basic Version - T4748

Measuring Range

Measuring Time

Accuracy / Notes

Test

Water-in-Oil

Measuring Range

0 – 1.0 vol.% H₂O

Measuring Time

2 – 20 min (depending on value)

Accuracy / Notes

±3%

Test

Base Number (BN)

Measuring Range

0 – 150 BN

Measuring Time

2 – 20 min (depending on value)

Accuracy / Notes

±1 BN

Test

Oil Viscosity

Measuring Range

Max. lube oil range

Measuring Time

Approx. 3 min

Accuracy / Notes

-

Test

Heated Electronic Falling Ball Viscosity

Measuring Range

Viscosity: 1 – 999 mm²/s Density: 0.82 – 1.05 g/ml

Measuring Time

Approx. 15 min

Accuracy / Notes

±3% (Sample: 200 ml)

Test

Saltwater Contamination

Measuring Range

Go / No-Go

Measuring Time

Approx. 10 min

Accuracy / Notes

-

Test

Spot Test

Measuring Range

All lube oils

Measuring Time

Approx. 1 min

Accuracy / Notes

-

Test

Compatibility Check

Measuring Range

All heavy fuel oils + MGO

Measuring Time

Approx. 10 min

Accuracy / Notes

Go / No-Go (Sample: 2 × 50 ml)

Test

Measuring Range

Measuring Time

Accuracy / Notes

Test

T4847 Option I also includes:

Measuring Range

Measuring Time

Accuracy / Notes

Test

Flash Point Check

Measuring Range

Ambient temperature – 200 °C

Measuring Time

Approx. 15 min

Accuracy / Notes

±2% (<100 °C), ±6 °C (>100 °C)

Test

Pour Point Check

Measuring Range

-20 °C to +30 °C

Measuring Time

Approx. 10 min

Accuracy / Notes

±2 °C

Test

Measuring Range

Measuring Time

Accuracy / Notes

Test

T4848 Option II also includes:

Measuring Range

Measuring Time

Accuracy / Notes

Test

MT Cat Fines Check

Measuring Range

-

Measuring Time

Approx. 15 min

Accuracy / Notes

Up to 8 simultaneous measurements, visually quantifiable

Test

Total Iron Check

Measuring Range

15/20 – 1100 mg/kg (ppm)

Measuring Time

Approx. 20 min (for two cylinder drain oil samples)

Accuracy / Notes

±20 mg/kg (ppm) at 70 °C using LED illuminance meter

Contact us for a FREE quote today

We are always available and our team is happy to assist
with any urgent issues.